Konica-minolta PULSOX 300 User Manual

Browse online or download User Manual for Health and hygiene Konica-minolta PULSOX 300. Konica Minolta PULSOX 300 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - OXYGEN SATURATION MONITOR

OXYGEN SATURATION MONITORPULSOX -300®-300®-300Instruction ManualThis instrument must be used according to the instructions of the doctor.

Page 2 - Safety Symbols

8ForewordNotes on Use• This instrument should be used under the following operat-ing conditions.• Temperature of 0 to 40°C (32 to 104°F), relative h

Page 3 - Safety Precautions

9ForewordDifference from the SaO2value to displayed value<Table 1> <Table 2>SaO2HbCOSaO2Methemoglobin1% 5% 10% 1% 5% 10%50 5070 7090 90100

Page 4

10ForewordNotes on Storage• This instrument should be stored under the following storage conditions.• Temperature of –10 to 60°C (14 to 140°F), rela

Page 5 - Contents

11ForewordDisposal Method• When disposing of the used battery, insulate the polarity ter-minals with insulating tape etc. If the terminals of the b

Page 6 - Foreword

12PreparationsPreparationsBefore starting measurements using this instrument, follow the procedure given below to ensure that measurements are stable.

Page 7 - About Probe

13Preparations4 Pass the other end of the band outward through band fi xture B.◆Take care not to twist the band.the band.2) Connecting the probeC

Page 8 - System Diagrams

14Preparations3) Installing the batteryProcedure1 Turn the battery cover clockwise approximately 90 degrees to open the cover.◆W h e n t u r n

Page 9 - Names of Parts

15MeasurementMeasurement1) Starting the Measurement1 Attach the probe to a fi nger of an adult patient.2 Press the Power button. ●The Power is ON

Page 10 - Notes on Use

16Measurement2) Reading the displayWhen the instrument is turned ON, all the elements of the LED will light up.The pulse level, pulse rate and Sp

Page 11

17Measurement3) Ending the MeasurementAbout Auto Power-Off functionThe instrument has an auto power-off function that operates independently. If no

Page 12 - Notes on Storage

Safety SymbolsWarnings and precautions noted in this manual and the instrument are indicated by the following markings, designed to prevent acc

Page 13 - Disposal Method

18Check PointsTrouble ShootingsError MessagesThe instrument display error messages when the error occurs. Check and correspond to them correctly.

Page 14 - Band fi xture A

19Check Points                        

Page 15 - Band fi xture B

20Check Points                                                

Page 16 - 3) Installing the battery

21Check Points                        

Page 17 - Measurement

22Check PointsMaintenance and Inspection• Before using the instrument, make sure that there is no dam-age to the instrument, no damage or wire-

Page 18 - 2) Reading the display

23AppendixSpecifi cation Dual-wavelength pulse type (665nm/880nm) arterial blood oxygen satura-tion measuring instrument • Name: Oxygen Saturation

Page 19

24AppendixProbes (SR-5C, SP-5C, LM-5C, SD-5C)Probes (SR-5C, SP-5C, LM-5C, SD-5C) LED-SPD probe Type• Finger Clip Probe SR-5C• Spot Check Probe SP-

Page 20 - Error Messages

25AppendixAppendixCalculation Method for Displayed Values<Pulse level meter>The pulse level meter displays the transmitted light amount ratio ca

Page 21 - Check Points

26AppendixEMC GuidanceGuidance and manufactures declaration - electromagnetic emissionsGuidance and manufactures declaration - electromagnetic immun

Page 22

27AppendixGuidance and manufactures declaration - electromagnetic immunityThe Oxygen Saturation Monitor Model PULSOX-300 is intended for use in the e

Page 23

1ForewordSafety PrecautionsTo ensure correct use of this instrument, read the following points carefully and adhere to them. After you have read this

Page 24 - Check Points before repairing

28Appendix Spectral Absorption of Hb and HbO2Absorption coefficientWavelengthRed light Infrared lightMeasurement PrincipleThis instrument measures th

Page 25 - Specifi cation

29AppendixRelation between Oxygen Saturation and Partial PressureThe relation between oxygen saturation (SpO2) and oxygen partial pressure (PaO

Page 26 - Appendix

AGGAKK (AGGAKK (AGGAKK1)Printed in Japan9222-1732-25© 2005-2006 KONICA MINOLTA SENSING, INC.Manufacturer3-91, Daisen-nishimachi, Sakai-ku, Sakai-shi,

Page 27

2ForewordCAUTION(Failure to adhere to the following points may result in injury or damage to the in-strument or other property.)strument or other prop

Page 28 - EMC Guidance

3ForewordContentsSafety SymbolsSafety Precautions ...1Foreword ...

Page 29

4ForewordForewordSafety NotesCAUTION: Federal law restricts this device to sale by or on the order of a physician.The instrument is designe

Page 30 - Red light Infrared light

5ForewordPackage ContentsBefore using the instrument, make sure that the following items are present.1) PULSOX-300 main body × 12) Wristband WB-300 ×

Page 31

6ForewordSystem Diagrams and AccessoriesThe instrument consists of the following items.Optional accessoriesStandard accessoriesWristband WB-

Page 32 - Manufacturer

7ForewordNames of PartsPULSOX-300Power buttonProbe connector WristbandBattery coverUsed to start/end measurement.Measurement results and messages are